Skip to main content
Üsküdar / İstanbul
Walchem

Walchem Amperometric Disinfection Sensors (Chlorine Dioxide, Peracetic Acid, Hydrogen Peroxide, Ozone, Chlorite)

Walchem amperometric sensors measure chlorine dioxide, peracetic acid, hydrogen peroxide, ozone and chlorite continuously without reagents, for W100, W600 and W900 controllers and the Intuition series.

Send us your requirement

Walchem Amperometric Disinfection Sensors (Chlorine Dioxide, Peracetic Acid, Hydrogen Peroxide, Ozone, Chlorite) — describe your application and we will help you identify the right model and configuration together.

Technical specifications

Two gray cylindrical Walchem amperometric disinfection sensors with membrane caps at the sensing end
Measuring principleContinuous, direct amperometric measurement of the chemical concentration without reagents; the sample can be returned to the process
Parameters in this brochureChlorine dioxide (ClO2), peracetic acid (PAA), hydrogen peroxide (H2O2), ozone (O3) and chlorite; Walchem also offers chlorine, bromine and stabilized bromine sensors
Measuring rangesClO2: 0-2 and 0-20 ppm; PAA: 0-200, 0-2000 and 0-20000 ppm; H2O2: 0-200, 0-2000 and 0-20000 ppm; O3: 0-2 and 0-20 ppm; chlorite: 0-2 ppm
ResolutionClO2 0.001 / 0.1 ppm; PAA and H2O2 0.1 / 1 / 10 ppm; O3 0.001 / 0.01 ppm; chlorite 0.001 ppm (by range)
CalibrationWeekly for all sensors; PAA with thiosulfate titration, ozone with DPD-4
Sample conditionsFlow 30-100 L/h; conductivity at least 10 µS/cm; sample pH: ClO2 1-12, PAA 1-6, H2O2 2-11, O3 4-9, chlorite 6-9
Response and conditioning timeClO2 1.5 min / 60 min; PAA 3 min / 3 hours; H2O2 8 min / 5 hours; O3 8 min / 60 min; chlorite 1 min / 24 hours
InterferencesClO2: ozone (2500%); PAA: ozone (2500%), ClO2 (100%), H2O2; H2O2: chlorine, PAA and ozone have no effect; O3: chlorine (1.5%), ClO2; chlorite: Mn2+, Fe2+, nitrate
Pressure and temperature1 bar for ClO2, PAA, H2O2 and O3 (chlorite 5 bar); temperature ClO2 0-50 °C, PAA 0-45 °C, H2O2 0-60 °C, O3 0-45 °C, chlorite 0-40 °C
Electrical±5 VDC, 10 mA maximum supply; signal 0 to -2000 mV DC; cable extension up to 30 m with a 2 twisted pair, 22 AWG shielded cable (Walchem 100084)
Materials and connectionsPVC and stainless steel (chlorite sensor: microporous hydrophilic membrane, PVC, PEEK, stainless steel); flow cell inlet 1/4 inch NPTF, outlet 3/4 inch NPTF
Flow cell optionsW600-DS-PN: flow switch manifold on panel, about 0.9 m cable; W600-DS-FN: flow cell, 6 m cable
Controller compatibilityThe brochure lists W100, W600 and W900 series controllers; disinfection sensor options also appear in the Intuition-6 and Intuition-9 brochures

Walchem amperometric disinfection sensors measure the concentration of an oxidizing disinfectant directly and continuously, without reagents and without discarding the sample. The sensor family covers chlorine dioxide, peracetic acid (PAA), hydrogen peroxide, ozone and chlorite, and Walchem also supplies chlorine, bromine and stabilized bromine sensors. Each sensor connects to a Walchem controller and sits in a flow cell. As the Walchem distributor in Turkey, Mikron Proses (Üsküdar, Istanbul) supplies the sensors, flow cells, membrane caps and electrolyte, and provides application consulting for sensor and controller selection, commissioning and calibration support.

Tell us the disinfectant and the expected concentration and we will select the sensor range → Send us your requirement

Selection table by parameter

ParameterRanges (ppm)Sample pHResponse / conditioningPressureMain interferences
Chlorine dioxide (ClO₂)0-2, 0-201-121.5 min / 60 min1 barOzone (2500%)
Peracetic acid (PAA)0-200, 0-2000, 0-200001-63 min / 3 hours1 barOzone (2500%), ClO₂ (100%), H₂O₂
Hydrogen peroxide (H₂O₂)0-200, 0-2000, 0-200002-118 min / 5 hours1 barNot affected by chlorine, PAA and ozone
Ozone (O₃)0-2, 0-204-98 min / 60 min1 barChlorine (1.5%), ClO₂
Chlorite0-26-91 min / 24 hours5 barMn²⁺, Fe²⁺, nitrate

Each parameter is offered in more than one range with separate part numbers, so the range can be matched to the expected concentration. A range that is wider than needed gives a coarser resolution at low concentrations. The sensors differ in their interference profiles: for example the hydrogen peroxide sensor is not affected by chlorine, PAA or ozone, while the ozone sensor responds slightly to chlorine (1.5%) and to chlorine dioxide.

Sample conditions and maintenance

All sensors need a steady sample flow of 30-100 L/h and a conductivity of at least 10 µS/cm. The permitted sample pH differs by sensor, from the wide range of the chlorine dioxide sensor (pH 1-12) to the narrow range of the PAA sensor (pH 1-6). Calibration is weekly for every sensor; PAA is calibrated by thiosulfate titration and ozone by the DPD-4 method. Conditioning time after installing a new sensor ranges from 60 minutes for chlorine dioxide and ozone to 24 hours for chlorite. Electrolyte stored in its original bottle at 5-35 °C, protected from sunlight, lasts one year or until its expiry date, and a sensor stored dry without electrolyte has an unlimited shelf life.

Controller connection

The sensors are powered with ±5 VDC (10 mA maximum) and give a signal from 0 to -2000 mV DC, which can be extended up to 30 m with a shielded 22 AWG cable (Walchem 100084). The brochure lists the W100, W600 and W900 series controllers, and the Intuition-6 and Intuition-9 brochures include disinfection sensor options as well; see the Intuition-6 series, the Intuition-9 series and the W100P panel-mount controller. Choose one sensor and one flow cell (panel flow switch manifold or flow cell). For related background see the hydrogen peroxide measurement guide, the pool chlorine measurement guide and the Walchem brand page.

Applications

  • Continuous disinfectant monitoring and control with chlorine, bromine, chlorine dioxide, peracetic acid, hydrogen peroxide, ozone and chlorite sensors

Datasheets and manuals

Documents were last compared with the manufacturer's source on 2026-10-30.

Send us your requirement for Walchem Amperometric Disinfection Sensors (Chlorine Dioxide, Peracetic Acid, Hydrogen Peroxide, Ozone, Chlorite)

Describe your application; we will help you identify the right model and configuration.

Call Quote