Skip to main content
Üsküdar / İstanbul
Hefer

Hefer WHA Series Online Water Hardness Analyzer (WHA-3000-C / WHA-2000)

Online water hardness (CaCO3) monitoring at softener outlets, RO pretreatment and boiler feed water with the Hefer WHA-3000-C analyzer (0.1-999 ppm) and the WHA-2000 ion-selective electrode (ISE) sensor.

Send us your requirement

Hefer WHA Series Online Water Hardness Analyzer (WHA-3000-C / WHA-2000) — describe your application and we will help you identify the right model and configuration together.

Technical specifications

Hefer WHA-3000-C hardness analyzer panel: a controller with an LCD above two electrodes in a clear sample cell, with a pressure gauge and a valve, inside a clear-door enclosure
ModelsWHA-3000-C complete panel analyzer (IC-3000 controller + sample cell) and WHA-2000 ISE hardness sensor
Measuring principleIon-selective electrode (ISE) potentiometric measurement of total hardness (Ca²⁺ / Mg²⁺)
WHA-3000-C measuring range0.1-999 ppm CaCO₃ (datasheet); IC-3000 controller display scale 0.01-1000 ppm (selectable 10.00 / 100.0 / 1000 ppm, autoranging)
WHA-2000 concentration range0.2 ppm Ca/Mg to saturated solution (technical notes); 0.4-40,000 ppm Ca²⁺ (sensor datasheet)
Resolution and accuracy (WHA-3000-C)0.01 / 0.1 ppm depending on the scale; accuracy ±5% of the reading
Slope and repeatability (WHA-2000)25 ± 5 mV between 10 and 100 ppm (25°C); repeatability ±2%; isopotential point 1 ppm Ca/Mg; resistance 100 MΩ
Response time (WHA-2000)95% of the reading within 30 seconds (above 10 ppm)
Temperature and pH rangeSample water 0-40°C (acrylic sample cell 0-50°C); pH 5-10
Pressure and sample flowWHA-3000-C process inlet pressure 0.5-10 bar; sample flow 10-100 L/h with 6 mm / 12 mm connections; WHA-2000 sensor working pressure 0-2 bar
Chemical limits (WHA-2000)Free chlorine below 0.1 ppm; sulfate (SO₄²⁻) below 50 ppm; carbonate / bicarbonate below 280 ppm
Signal and relay outputsIsolated 4-20 mA (or 0-20 mA) analog output; 2 × SPDT (5 A) dry-contact relays (the controller provides 4 dry contacts and an optional RS-232)
Enclosure and supplyIP-65 enclosure with a clear door (350 × 350 × 180 mm); 110/220 VAC ±10%, 50/60 Hz

The Hefer WHA series is an online water hardness analyzer family: the WHA-3000-C is a complete panel analyzer for 0.1-999 ppm CaCO₃ and the WHA-2000 is the ion-selective electrode (ISE) hardness sensor. The analyzer is designed for potable water, drinking water and water treatment applications such as softeners and reverse osmosis (RO) systems, and it sends the reading through an isolated 4-20 mA output. As the Hefer distributor in Turkey, Mikron Proses (Üsküdar, Istanbul) handles selection, supply, commissioning, field calibration and service of WHA-3000-C analyzer panels and replacement WHA-2000 ISE sensors.

By monitoring total water hardness (CaCO₃) in real time at the outlet of a softener, in RO pretreatment lines and in boiler feed water, the analyzer is meant to catch the moment a softener resin reaches saturation and starts to leak hardness, which protects RO membranes and boiler tubes against scaling.

WHA-3000-C analyzer and WHA-2000 sensor compared

The table separates the parameters of the complete analyzer panel (WHA-3000-C) from those of the sensor element (WHA-2000), as given in the manufacturer’s datasheets and manuals.

ParameterWHA-3000-C complete online analyzer (IC-3000 panel)WHA-2000 hardness sensor (replacement / OEM electrode)
Measuring range (datasheet)0.1 - 999 ppm CaCO₃0.4 - 40,000 ppm Ca²⁺ (datasheet) / 0.2 ppm Ca/Mg to saturated solution (technical notes)
Controller display scale0.01 - 1000 ppm (10.00 / 100.0 / 1000 ppm autoranging)Depends on the connected controller (IC-3000 or a compatible controller with mV input)
Resolution, accuracy, repeatabilityResolution 0.01 / 0.1 ppm; accuracy ±5% of the readingRepeatability ±2%; slope 25 ± 5 mV between 10 and 100 ppm (25°C)
Response time, isopotential point-95% of the reading in 30 seconds (above 10 ppm); isopotential point 1 ppm Ca/Mg
Working pressureProcess inlet 0.5 - 10 bar0 - 2 bar at the sensor
Sample temperature and pHWater 0 - 40°C (acrylic cell 0 - 50°C)0 - 40°C; pH 5 - 10 (a 1°C temperature change can cause a deviation of up to 2%)
Sample flow and connections10 - 100 L/h adjustable acrylic sample cell (6 mm inlet / 12 mm outlet)12 mm diameter, 110 mm length; needs a steady flow without air bubbles
Outputs and enclosureIsolated 4-20 mA (or 0-20 mA), dry-contact relays, IP-65 (350 × 350 × 180 mm)mV signal with 100 MΩ resistance (to the controller input)
Sensor life-About 1 - 2 years, depending on use and water conditions

Sources: hefer-wha-3000c-datasheet.pdf, hefer-wha-3000c-kullanim-kilavuzu.pdf, hefer-wha-2000-sertlik-sensoru-datasheet.pdf and hefer-wha-2000-sertlik-sensoru-aciklamalar-kilavuzu.pdf.

Sample conditions, chemical limits and field calibration

For the ISE membrane to work accurately and last, the sample water has to stay within these limits:

  1. Oxidizing agents and organics: use the sensor in aqueous solutions only. Free chlorine, ozone and organic solvents can damage the sensor membrane permanently, and free chlorine in the sample must be below 0.1 ppm.
  2. Complex formation: sulfate, bicarbonate and carbonate ions form complexes with free calcium and magnesium and can shift the reading. Sulfate (SO₄²⁻) should be below 50 ppm and carbonate / bicarbonate below 280 ppm.
  3. Cation interference (10% deviation thresholds, WHA-2000 technical notes, Table 1): at a total hardness of 10 ppm CaCO₃, the concentrations that cause a 10% deviation are Cu²⁺ 1.9 ppm, Zn²⁺ 1.95 ppm, Fe²⁺ 3.3 ppm, Ni²⁺ 5.7 ppm, Sr²⁺ 35 ppm, Ba²⁺ 82.2 ppm, Na⁺ 660 ppm and K⁺ 3,909 ppm. If a deviation appears after exposure to such ions, the sensor can be rested in deionized water; long exposure can cause irreversible damage.
  4. Installation and routine calibration:
    • Installation calibration (2-point program): take one sample of raw water (before softening) and one from the softener or RO outlet, check their hardness with a hardness test, and enter the values. For higher accuracy, 2 to 5 points can be taken from 0.10 / 1 / 10.00 / 100.0 / 1000 ppm standard solutions.
    • Routine field calibration (one point): a 3-way valve on the acrylic sample cell supplies soft and hard water, and a one-point reference correction takes no longer than 2 minutes.

When the response time becomes much longer or the calibration slope (25 ± 5 mV between 10 and 100 ppm at 25°C) drops, the WHA-2000 sensor is replaced.

Applications

  • Monitoring hardness leakage at ion-exchange water softener outlets and triggering regeneration
  • Membrane scaling protection in reverse osmosis (RO) and electrodeionization (EDI) pretreatment lines
  • Continuous hardness control in steam boiler feed water and cooling tower make-up lines
  • Blended hardness control in drinking water and process water preparation plants

Send us your requirement for Hefer WHA Series Online Water Hardness Analyzer (WHA-3000-C / WHA-2000)

Describe your application; we will help you identify the right model and configuration.

Call Quote