Hamilton Calibration Solutions (DuraCal pH and ORP Buffers and Conductivity Standards)

Hamilton DuraCal pH and ORP buffer solutions and conductivity calibration standards for field calibration of process and laboratory sensors, traceable to NIST and PTB and verified by DAkkS and DFM.

Send us your requirement

Hamilton Calibration Solutions (DuraCal pH and ORP Buffers and Conductivity Standards) — describe your application and we will help you identify the right model and configuration together.

Technical specifications

Hamilton DuraCal pH, ORP and conductivity calibration solutions
pH buffer seriesNominal pH 1.09, 1.68, 2.00, 3.06, 4.01, 5.00, 6.00, 7.00, 8.00, 9.21, 10.01, 11.00 and 12.00 (13 values in the 2024 catalog); the 2026 catalog adds pH 9.00
pH accuracyUp to ±0.01 pH at pH 4.01 and 7.00; ±0.02 pH for most other values; ±0.05 pH at pH 12.00
pH stabilityUp to 60 months after the date of manufacturing; the pH 4.01 / 7.00 / 10.01 set (REF 238924) is specified as ±0.01 pH for 24 months or ±0.02 pH for 60 months
pH traceabilityClosed-loop (top-down and bottom-up) traceability to NIST (USA) and PTB (Germany); a DAkkS calibration certificate for every DuraCal production batch
pH packaging250 mL and 500 mL bottles with a calibration compartment, 3 x 500 mL sets, 5 L and 10 L; about 15 mL of buffer is used per calibration
Multi-buffer calibration setspH 4.01 / 7.00 / 9.21 (REF 238922) and pH 4.01 / 7.00 / 10.01 (REF 238924), each as 3 x 500 mL
ORP buffers271 mV and 475 mV nominal; ±5 mV accuracy; 24 months stability; 250 mL and 500 mL packages
Conductivity standards range1.3 µS/cm to 100 mS/cm: 1.3, 5, 15, 84, 100, 147, 706, 1413 and 12,880 µS/cm and 100 mS/cm
Conductivity standard accuracyPatented 1.3 and 5 µS/cm standards with ±1% certified accuracy; the procedure for determining conductivity was developed in collaboration with DFM (Denmark) and checked by PTB
Conductivity standard stability and handling12 months for 1.3 µS/cm and up to 36 months for the other values depending on bottle and value (18 months for the 500 mL Calpack bottles); bottles may stay open for a total of 60 minutes
Bottle designSqueeze-to-fill bottle with an integrated calibration compartment; the remaining main volume stays closed and uncontaminated

Hamilton DuraCal pH and ORP buffer solutions and conductivity standards are certified reference solutions for the field calibration of process and laboratory sensors. The pH buffers are traceable to NIST (USA) and PTB (Germany) through closed-loop traceability, and the conductivity standards are certified with DFM. As one of Hamilton’s distributors in Turkey, Mikron Proses (Üsküdar, Istanbul) supplies calibration solutions together with Hamilton sensors and provides application consulting on their selection.

Need pH, ORP or conductivity calibration solutions? → Send us your requirement

The complete value tables with order codes are on the DuraCal pH buffer solutions page. The sensors to be calibrated are described in Hamilton pH and ORP process sensors, the Hamilton conductivity sensors page and the bioprocess pH sensors page.

pH and ORP buffers: closed-loop traceability

The value of a DuraCal pH buffer comes from Hamilton’s closed-loop approach. Top-down, the buffer value is determined against secondary reference solutions that are compared against primary solutions from NIST or PTB. Bottom-up, a representative number of samples from every production lot is verified again by an independent DAkkS laboratory, and every production batch receives its own DAkkS calibration certificate.

The pH values run from 1.09 to 12.00. Accuracy is up to ±0.01 pH at pH 4.01 and 7.00 and ±0.02 pH for most other values. Stability is up to 60 months after manufacturing. For three-point calibration, 3 x 500 mL sets with pH 4.01 / 7.00 / 10.01 (REF 238924) or pH 4.01 / 7.00 / 9.21 (REF 238922) are available. For ORP (redox) measurement there are two buffers, 271 mV and 475 mV, with ±5 mV accuracy and 24 months stability.

The bottle has an integrated calibration compartment: you squeeze the bottle to fill the compartment, calibrate, and empty the compartment, so the remaining solution stays closed and uncontaminated. Only about 15 mL of buffer is used per calibration.

Conductivity standards: patented low-range stability

A conductivity standard is not a buffer, so for low values CO2 absorption or contamination can easily disturb the measurement. Hamilton states that it is the first manufacturer to offer patented standards of 1.3 and 5 µS/cm with a certified accuracy of ±1%; the determination procedure was developed in collaboration with DFM, the Danish institute for fundamental metrology, and confirmed by comparison measurements with PTB.

The range covers 1.3 µS/cm to 100 mS/cm. The 1.3 µS/cm standard is stable for 12 months and the other values for up to 36 months, depending on value and bottle. A bottle may stay open for a total of 60 minutes. For pure water background, see the pure water conductivity and USP 645 guide.

Compliance and use

The values are verified in ISO 17025 accredited laboratories (DAkkS and DANAK), and marked items in the catalog are ISO 17034 certified reference materials. These products are used where traceable calibration records are needed, for example in pharmaceutical, biopharma, food and beverage production.


Which pH, ORP and conductivity values you need depends on the sensor type, the calibration frequency and your regulatory requirements. Tell us about your sensors and we will define the right solution set together.

Applications

  • Periodic field calibration of process pH and ORP sensors (EasyFerm, Polilyte, ChemoTrode)
  • Calibration of laboratory benchtop pH meter electrodes
  • Verification of Conducell conductivity sensors in the low and high conductivity range
  • Regulated production environments that need traceable calibration records under GMP, GLP and ISO 9001
  • Verification of ORP (redox) measuring points