Hamilton ChemoTrode pH Sensors for Aggressive Media
Hamilton ChemoTrode pH sensors for aggressive chemical media: refillable liquid electrolyte, Everef-F reference cartridge, 0-14 pH, 0-130 °C and 6 bar g, with ChemoTrode P, Bridge and ORP variants.
Send us your requirementHamilton ChemoTrode pH Sensors for Aggressive Media — describe your application and we will help you identify the right model and configuration together.
Technical specifications
| Measuring range | 0 to 14 pH (recommended range 0 to 13 pH) |
|---|---|
| Process temperature | 0 to 130 °C |
| Pressure range | 0 to 6 bar g (relative to ambient) |
| Sterilization and cleaning | SIP and CIP |
| Electrolyte | Refillable liquid electrolyte, pressurizable up to 6 bar through a port in the sensor; ChemoTrode: viscous 3 M KCl-LR, ChemoTrode P: Protelyte |
| Reference system | Everef-F reference cartridge, which keeps the reference electrolyte free of silver and protein precipitation |
| pH glass and diaphragm | PHI glass; HP ceramic diaphragm |
| Shaft lengths | 120, 150, 200 and 250 mm (ChemoTrode P: 120, 150 and 250 mm) |
| Connectors and temperature sensor | S7, or VP6 with Pt1000 (Pt100 on the 250 mm ChemoTrode) |
| Variants | ChemoTrode, ChemoTrode P (Protelyte), ChemoTrode Bridge (Everef-B reference, Skylyte electrolyte) and ChemoTrode ORP |
Hamilton ChemoTrode is a pH sensor with a refillable liquid electrolyte for demanding applications in the chemical and pharmaceutical industries. It measures 0 to 14 pH at 0 to 130 °C and up to 6 bar g, and it can be sterilized and cleaned with SIP and CIP. As one of Hamilton’s distributors in Turkey, Mikron Proses (Üsküdar, Istanbul) supplies ChemoTrode sensors and provides application consulting on sensor selection for aggressive media.
Need a pH sensor for an aggressive process? → Send us your requirement
For the other Hamilton pH and ORP sensor families, see Hamilton pH and ORP process sensors; for general selection advice, see the pH electrode selection guide.
How the ChemoTrode is built
The Everef-F reference cartridge keeps the reference electrolyte free of silver and prevents the precipitation of proteins, which extends the electrode lifetime by preventing diaphragm clogging. The liquid electrolyte is refilled and pressurized through a port in the sensor, up to 6 bar. Hamilton lists fast response time and high measurement precision as benefits of the liquid electrolyte.
ChemoTrode variants
| Variant | Measuring range | Reference and electrolyte | Temperature and pressure |
|---|---|---|---|
| ChemoTrode | 0 to 14 pH | Everef-F, viscous 3 M KCl-LR | 0 to 130 °C, 6 bar g |
| ChemoTrode P | 0 to 14 pH | Everef-F, Protelyte | 0 to 130 °C, 6 bar g |
| ChemoTrode Bridge | 0 to 14 pH | Everef-B, Skylyte | 0 to 130 °C, 6 bar g |
| ChemoTrode ORP | ± 2000 mV (platinum ring) | Everef-F, viscous 3 M KCl-LR | 0 to 130 °C, 6 bar g, CIP and SIP |
All variants use an HP ceramic diaphragm. The ORP version also has a refill hole for electrolyte refilling and pressurization of the reference electrolyte.
Which ChemoTrode variant and shaft length fit your process depends on the medium, the temperature and the installation. Tell us about your application and we will help you choose.
Applications
- Acid and base neutralization processes
- Chemical and petrochemical reactors
- Electroplating and metal surface treatment
- Aggressive wastewater treatment
Datasheets and manuals
Other Hamilton products
- Hamilton Arc Peripherals and ArcAir Software (Arc View Mobile, Arc Wi Pivot, Arc USB, ArcAir) Online Measurement and Automation Systems
- Hamilton CO2NTROL Solid-State Optical Dissolved CO2 Sensor Online Measurement and Automation Systems
- Hamilton DuraCal pH Buffer, ORP and Conductivity Calibration Solutions Online Measurement and Automation Systems
- Hamilton EasyControl pH and ORP Sensors pH and ORP Sensors
- Hamilton Bioprocess pH Sensors (EasyFerm Plus, EasyFerm Bio and OneFerm) pH and ORP Sensors
- Hamilton GlucoSense In-Situ Optical Glucose Sensor for Bioreactors Online Measurement and Automation Systems
- Hamilton Cell Density Sensors (InCyte / Dencytee) Online Measurement and Automation Systems
- Hamilton Conductivity Sensors Conductivity Sensors
- Hamilton InchTrode pH Electrode pH and ORP Sensors
- Hamilton Incyte Arc Viable Cell Density Sensor Online Measurement and Automation Systems
- Hamilton IonoTrode pH Electrode for Low-Conductivity Media pH and ORP Sensors
- Hamilton Calibration Solutions (DuraCal pH and ORP Buffers and Conductivity Standards) Online Measurement and Automation Systems
- Hamilton Laboratory pH and ORP Electrodes (GLP Series) Online Measurement and Automation Systems
- Hamilton Process Analytics, Liquid Handling and Laboratory Solutions Online Measurement and Automation Systems
- Hamilton Microlab 600 and 700 Diluters, Dispensers and Syringe Pumps Online Measurement and Automation Systems
- Hamilton MecoTrode pH Electrode Hamilton pH and ORP Process Sensors
- Hamilton OEM Sensors and Sensor Accessories Online Measurement and Automation Systems
- Hamilton Oxygen Sensor Accessories (Oxylyte Electrolyte, Membrane Kits, Calibration Station) Hamilton Dissolved Oxygen Sensors
- Hamilton OneFerm pH Single-Use pH Sensor Hamilton pH and ORP Process Sensors
- Hamilton Polilyte Plus pH Sensors Hamilton pH and ORP Process Sensors
- Hamilton Process Fittings and Flow Cells (FlexiFit, StatoFlex, FlowChamber, FlowCell, FlexiFlow, MasterFit) Online Measurement and Automation Systems
- Hamilton Retractex Retractable Sensor Housings Online Measurement and Automation Systems
- Hamilton Precision Syringes, Needles and Chromatography Accessories Online Measurement and Automation Systems
- Hamilton VisiFerm DO, OxyFerm, OxyGold and Oxysens Dissolved Oxygen Sensors Hamilton Dissolved Oxygen Sensors
Send us your requirement for Hamilton ChemoTrode pH Sensors for Aggressive Media
Describe your application; we will help you identify the right model and configuration.