Hamilton Laboratory pH and ORP Electrodes (GLP Series)
Hamilton GLP laboratory pH and ORP electrodes, each supplied with an indelible serial number and an individual test certificate. The range includes Polilyte Lab, Single Pore Glass, BioTrode, FoodTrode, SlimTrode and Liq-Glass electrodes, S7 cables, electrolytes and DuraCal buffers. Mikron Proses is one of Hamilton's distributors in Turkey.
Send us your requirementHamilton Laboratory pH and ORP Electrodes (GLP Series) — describe your application and we will help you identify the right model and configuration together.
Technical specifications

| Measuring range | pH 0 to 14; ORP models ±2000 mV |
|---|---|
| Temperature range by model | Liq-Glass and FoodTrode -10 to 100 °C; Polilyte Lab and FlushTrode -10 to 80 °C; Gel-Glass -10 to 60 °C; Single Pore Glass, BioTrode, SlimTrode, MiniTrode, SpinTrode and TipTrode 0 to 100 °C; Polyplast, Double Pore, FillTrode and FlaTrode 0 to 60 °C |
| Diaphragm technologies | Single Pore (a capillary pore about 200 times larger in cross-section than the pores of a ceramic diaphragm), Double Pore (2 Single Pores), ceramic diaphragms (1 or 3), ring diaphragm and sleeve diaphragm (FlushTrode) |
| Reference systems and electrolytes | Everef and Everef-B reference systems; Polisolve polymer electrolyte (acrylamide free); Protelyte, Skylyte-CL and 3 M KCl |
| Small sample volumes | BioTrode and MiniTrode with a 3 mm shaft and an immersion depth of only 7 mm, for very small volumes such as vials and microtiter plates; FlaTrode with a flat membrane for surfaces and single drops |
| Connector and IP rating | S7 connector head with a high-quality seal between electrode head and cable (IP 68), or fixed cables with BNC, DIN or Lemo plugs and a Pt1000 or NTC 30 kΩ temperature sensor |
| GLP traceability | Each electrode carries an indelible serial number and comes with an individual test certificate that lists the measured millivolt values |
Hamilton GLP laboratory pH and ORP electrodes are combination electrodes for daily laboratory work in GLP environments. Every electrode has an indelible serial number printed on the shaft and comes with an individual test certificate that lists the measured millivolt values, which supports traceability in your documentation. Mikron Proses (Üsküdar, Istanbul) is one of Hamilton’s distributors in Turkey. We supply laboratory electrodes, S7 cables, electrolytes and DuraCal pH buffers, and we provide application consulting on matching an electrode to your benchtop pH meter and sample type.
Laboratory electrodes · S7 cables · Electrolytes → Send us your requirement
Electrode selection by sample type
The model codes (REF) and technical limits below are taken from the Hamilton GLP Laboratory Electrodes Catalog (electrode pages and the specifications table).
| Application group | Model and main REF | Electrolyte and reference system | Diaphragm and shaft | Temperature and typical samples |
|---|---|---|---|---|
| Ion-weak, organic and difficult samples | Polilyte Lab (REF 238403; temperature models 242058, 242059, 242060, 242062) | Polisolve (maintenance-free); Everef-B | Single Pore; glass shaft | -10 to 80 °C; emulsions, suspensions, ion-weak media |
| General high-accuracy laboratory work | Single Pore Glass (REF 238160) | Skylyte-CL (refillable, REF 242080); Everef | Single Pore; glass shaft | 0 to 100 °C; emulsions, ion-weak media, general laboratory applications |
| Protein and very small volumes | BioTrode (REF 238140) | Protelyte (refillable, REF 238038); Everef | Ceramic diaphragm; 3 mm shaft, 7 mm immersion depth | 0 to 100 °C; solutions containing protein, microtiter plates |
| Test tubes, vials and NMR tubes | SlimTrode (REF 238150), MiniTrode (REF 238100), SpinTrode (REF 238197) | 3 M KCl (refillable); Everef | Ceramic diaphragm; shaft below 6 mm (SlimTrode), 3 mm (MiniTrode, SpinTrode) | 0 to 100 °C; test tubes (SlimTrode), vials (MiniTrode), NMR tubes (SpinTrode) |
| Foodstuffs with protein | FoodTrode (REF 238285) | Protelyte (refillable); Everef | 3 ceramic diaphragms; glass shaft | -10 to 100 °C; media containing proteins |
| Solid and semisolid samples | TipTrode (REF 238080); Double Pore (REF 238400) | TipTrode: Protelyte, Everef. Double Pore: Polisolve, Ag/AgCl | TipTrode: ceramic diaphragm. Double Pore: 2 Single Pores | TipTrode 0 to 100 °C; Double Pore 0 to 60 °C; puncture electrodes, Double Pore for cheese and meat |
| Flat surfaces | FlaTrode (REF 238401) | Polisolve (maintenance-free); Everef | Ring diaphragm; flat glass membrane; plastic shaft | 0 to 60 °C; surfaces such as paper and agar plates |
| Viscous samples and very low conductivity | FlushTrode (REF 238060) | 3 M KCl (refillable, REF 238036); Everef | Sleeve (ground sleeve) diaphragm; glass shaft | -10 to 80 °C; viscous samples, ion-weak media, cosmetics |
| Standard general laboratory use | Liq-Glass (REF 238000), Gel-Glass (REF 238025) | Liq-Glass: 3 M KCl (refillable); Everef. Gel-Glass: gel; Ag/AgCl | Ceramic diaphragm; glass shaft | Liq-Glass -10 to 100 °C; Gel-Glass -10 to 60 °C; routine aqueous solutions |
| Plastic shaft (portable and water applications) | Polyplast (REF 238380; BNC 238381; temperature models 242050, 242051, 242052) and FillTrode (REF 242064) | Polisolve (maintenance-free); Ag/AgCl (Polyplast), Everef (FillTrode) | Single Pore (Polyplast), ring diaphragm (FillTrode); plastic shaft | 0 to 60 °C; water and sewage |
| Laboratory ORP (redox) | Liq-Glass ORP (REF 238145), Polyplast ORP (REF 238385; BNC 238384) | Liq-Glass ORP: 3 M KCl; Everef. Polyplast ORP: Polisolve; Ag/AgCl | Platinum; 3 ceramic diaphragms (Liq-Glass ORP) or Single Pore (Polyplast ORP) | ±2000 mV; Liq-Glass ORP -10 to 100 °C, Polyplast ORP 0 to 60 °C |
For samples containing protein, the catalog recommends replacing the 3 M KCl in the refillable electrodes with Protelyte (REF 238038).
S7 cables, electrolytes and care solutions
Hamilton electrodes with an S7 head are connected to the pH meter with a 3 mm coaxial S7 cable. The equipment-side plug is chosen to fit the pH meter. All BNC plugs have a moveable protective cover that protects the plug from fluid splashes.
| Product group | Description and length or volume | Hamilton order code (REF) | Used with |
|---|---|---|---|
| S7 / BNC cable | 1 m, 3 m or 5 m | REF 355173 (1 m), 355176 (3 m), 355178 (5 m) | pH/ORP meters with a BNC input |
| S7 / DIN cable | 1 m, 3 m or 5 m | REF 355174 (1 m), 355177 (3 m), 355179 (5 m) | pH/ORP meters with a DIN input |
| S7 / Lemo cable | 1 m | REF 355175 | pH/ORP meters with a Lemo input |
| 3 M KCl electrolyte | 100 mL or 500 mL | REF 238036 (100 mL), 238936 (500 mL) | Liq-Glass, FlushTrode, SlimTrode, MiniTrode, SpinTrode, Liq-Glass ORP |
| Skylyte-CL electrolyte | 100 mL | REF 242080 | Single Pore Glass |
| Protelyte electrolyte | 100 mL | REF 238038 | BioTrode, FoodTrode, TipTrode |
| Storage solution | 500 mL, acid-buffered; also regenerates the electrode | REF 238931 | Storage of the electrodes |
| Cleaning solution set | Cleaning Solution A, Cleaning Solution B and storage solution; 15 to 30 minutes in each solution | REF 238290 | Removing contamination from the pH glass and the diaphragm |
Related pages
Applications
- pH and ORP analysis in GLP laboratory environments
- Foodstuffs: media containing proteins (FoodTrode) and solid and semisolid samples such as cheese and meat (TipTrode, Double Pore)
- Biochemistry, protein-containing solutions and very small sample volumes in vials, test tubes and microtiter plates (BioTrode, SlimTrode, MiniTrode, SpinTrode)
- Ion-weak media, emulsions, suspensions and viscous samples (Polilyte Lab, Single Pore Glass, FlushTrode)
Datasheets and manuals
- Hamilton GLP Laboratory pH and ORP Electrodes Catalog Catalog · PDF
- Hamilton DuraCal pH Buffer Solution Datasheet Datasheet · PDF
Documents were last compared with the manufacturer's source on 2026-10-30.
Other Hamilton products
- Hamilton Arc Peripherals and ArcAir Software (Arc View Mobile, Arc Wi Pivot, Arc USB, ArcAir) Online Measurement and Automation Systems
- Hamilton ChemoTrode pH Sensors for Aggressive Media pH and ORP Sensors
- Hamilton CO2NTROL Solid-State Optical Dissolved CO2 Sensor Online Measurement and Automation Systems
- Hamilton DuraCal pH Buffer, ORP and Conductivity Calibration Solutions Online Measurement and Automation Systems
- Hamilton EasyControl pH and ORP Sensors pH and ORP Sensors
- Hamilton Bioprocess pH Sensors (EasyFerm Plus, EasyFerm Bio and OneFerm) pH and ORP Sensors
- Hamilton GlucoSense In-Situ Optical Glucose Sensor for Bioreactors Online Measurement and Automation Systems
- Hamilton Cell Density Sensors (InCyte / Dencytee) Online Measurement and Automation Systems
- Hamilton Conductivity Sensors Conductivity Sensors
- Hamilton InchTrode pH Electrode pH and ORP Sensors
- Hamilton Incyte Arc Viable Cell Density Sensor Online Measurement and Automation Systems
- Hamilton IonoTrode pH Electrode for Low-Conductivity Media pH and ORP Sensors
- Hamilton Calibration Solutions (DuraCal pH and ORP Buffers and Conductivity Standards) Online Measurement and Automation Systems
- Hamilton Process Analytics, Liquid Handling and Laboratory Solutions Online Measurement and Automation Systems
- Hamilton Microlab 600 and 700 Diluters, Dispensers and Syringe Pumps Online Measurement and Automation Systems
- Hamilton MecoTrode pH Electrode Hamilton pH and ORP Process Sensors
- Hamilton OEM Sensors and Sensor Accessories Online Measurement and Automation Systems
- Hamilton Oxygen Sensor Accessories (Oxylyte Electrolyte, Membrane Kits, Calibration Station) Hamilton Dissolved Oxygen Sensors
- Hamilton OneFerm pH Single-Use pH Sensor Hamilton pH and ORP Process Sensors
- Hamilton Polilyte Plus pH Sensors Hamilton pH and ORP Process Sensors
- Hamilton Process Fittings and Flow Cells (FlexiFit, StatoFlex, FlowChamber, FlowCell, FlexiFlow, MasterFit) Online Measurement and Automation Systems
- Hamilton Retractex Retractable Sensor Housings Online Measurement and Automation Systems
- Hamilton Precision Syringes, Needles and Chromatography Accessories Online Measurement and Automation Systems
- Hamilton VisiFerm DO, OxyFerm, OxyGold and Oxysens Dissolved Oxygen Sensors Hamilton Dissolved Oxygen Sensors
Send us your requirement for Hamilton Laboratory pH and ORP Electrodes (GLP Series)
Describe your application; we will help you identify the right model and configuration.